This is the third step of the calibration phase. The patient must perform a flexion-extension movement with the assessed leg.
For this step, the KneeKG-3D–Tracker-FemoralTM and the KneeKG-3D-TibialTM must be GREEN.
If you have issue with this step, here’s a few tips:
Make sure all the needed trackers (KneeKG-3D–Tracker-FemoralTM and the KneeKG-3D-TibialTM) are in the field of view of the NDI camera or that the visibility is not obstructed.
If the patient is tall, you may have tilted up the camera, in the hip joint center calibration, to ensure the visibility of the pelvic tracker. Adjust back the camera, by tilting it down, so the needed trackers are visible.
For this movement, the patient must be parallel to the treadmill.
IMPORTANT: Before starting the recording, ensure the patient is holding their knee steady, not moving, in full extension and slightly forward.
Contact us by: